SPESIFIKASI
|
-
Merk : CORELAB / EA 206S |
|
|
- Buatan :
CHINA |
|
|
|
Features : |
|
|
|
Fully
automatic instruments : |
|
|
- The automatic sample tray is integrated with
the instrument, not separate structure, which is not only |
make the instrument elegant nice, but also
ensure the instruments reliability and security |
|
Large touch -
screen : |
|
|
|
-
5,7 inches large touch-screen LCD shows characters clearly and definitely |
|
- menu, working condition, operation prompt,
system parameters and test data on screen is clearly and definitely |
-
The man-machine conversation is completed by touching the corresponding
button on screen |
|
Thus brings users much convenience |
|
|
Strict
Safety precautions : |
|
|
-
The flow sensor can judge the working
conditions of pipeline at any time |
|
- The electron lock can lock the test stage
during testing to prevent error operation, avoid damaging the sample probe |
- The cuvette sensor can judge cuvette's
position, so it is unnecessary to set cuvette from the first position |
-
When stat chamber door is open, sample stage will be stopped, ensuring
operation safety |
|
Electrode
with superior performance : |
|
|
- The imported selective electrodes have
reliable quality and have a life-span of more than one year |
- At user's special demand, we can also supply
Natrium electrode. It's CV is less than 0.3% and its life-span |
is more than two years and a half |
|
|
Powerful
function : |
|
|
|
- Besides normal testing function, the
instrument also has functions of self-checking, precision testing, quality
control, |
cleaning, data output, operation prompt and sound prompt and so forth.
Among which, the self-checking function can monitor each operation units |
and check out different abnormal conditions
and then display fault states |
|
Automatic
calibration functions : |
|
|
- The special quality control calibration
function can calibrate standard curve by fixed-value serum, which guarantees
accuracy of test results |
No cross
contamination : |
|
|
|
- Reasonably scientific design of pipeline
makes cleaning automatically through the entry process, avoiding the cross
contamination between samples |
-
As sample flows through electrodes directly, no clogging can be caused |
|
Huge Storage
: |
|
|
|
-
1000 records can be stored for users to review and print at any time |
|
Other
Functions : |
|
|
|
- The device can be on all day long, and it
will be at standby states if no operation on it after ten minutes; Calibrate
one time in every 24 hours, saving reagents |
-
Electrode condition can be followed automatically, which frees the effect of
electrode floating |
|
-
One to three quality controls can be selected. Each control stores 30 quality
control results |
|
-
Seven parameters of five items of K, Na, Cl, iCa, nCa, Ph can be tested at
the same time |
|
Technical
Specifications : |
|
|
|
- Principle |
Ion selective electrode |
|
|
- Test Items |
K, Na, Cl, Ca, Ph |
|
|
- Sample |
Whole blood, serum,
plasma, urine and others (dialysate and etc.) |
|
- Throughput |
60 samples / hour |
|
|
- Sample
volume |
150
µl (when the flow sensor is set off, sample volume will
be consumed 200µl) |
- Sample
format |
Auto - mode 29 samples |
|
|
|
Stat mode 1 samples |
|
|
- Calibration |
Two-point
calibration. In whole blood mode, standard curve can be calibrated |
|
by quality control serum
with definite value |
|
- Output |
RS 232 standard port |
|
|
- Display |
5,7 inches LCD |
|
|
- Operation |
Touch screen |
|
|
- Printer |
Thermal
printer with high precision dot matrix, paper wide 110mm or EPSON stylus
printer |
- Working
conditions |
Temperature 10 ~ 30º C,
humility less than 70% |
|
- Power
supply |
220V; 50 Hz |
|
|
- Power
dissipation |
50 VA |
|
|
- Dimensions |
280 (W) x 340 (D) x 400
(H) [mm] |
|
- Weight |
About 18 [kg] |
|
|
- Performance
index : |
|
|
|
Item |
Test Range |
Resolation |
Repeatability |
K |
0.50 ~ 10.0 mmol/L |
0.01 mmol/L |
CV ≤ 1.0% |
Na |
20.0 ~ 200.0 mmol/L |
0.1 mmol/L |
CV ≤ 1.0% |
Cl |
20.0 ~ 200.0 mmol/L |
0.1 mmol/L |
CV ≤ 1.0% |
Ca |
0.20 ~ 5.00 mmol/L |
0.01 mmol/L |
CV ≤ 1.5% |
Ph |
4 ~ 9 Ph |
0.01 Ph |
CV ≤ 0.5% |
|
|
|
|